Send Inquiry Send SMS

PORTABLE PARTICLE COUNTER

PORTABLE PARTICLE COUNTER
PORTABLE PARTICLE COUNTER
Send Inquiry
Product Code : 9310,9510
Product Description

PORTABLE PARTICLE COUNTER:

SPECIFICATION: 

Size Range

9310-02 9510-02

9310: 0.3 to 25 μm 9510: 0.5 to 25 μm

Particle Channel Sizes

9310-02: 0.3, 0.5, 1.0, 3.0, 9510-02: 0.5, 0.7, 1.0, 3.0, 5.0, 10 μm, additional channel 5.0, 10 μm, additional channel

sizes available sizes available

Size Resolution

9310-02: <15% @ 0.5 μm 9510-02: <15% @ 0.5 μm

(per ISO 21501-4) (per ISO 21501-4)

Counting Efficiency

9310-02: 50% at 0.3 μm; 9510-02: 50% at 0.5 μm;

100% for particles >0.45 μm 100% for particles 0.75 μm

(per ISO 21501-4 and JIS) (per ISO 21501-4 and JIS)

Concentration Limit

820,000 particles/ft3 (29,000,000/m3) @ 10% coincidence loss

Light Source

Long life laser diode

Zero Count

<1 count per 5 minutes (per ISO 21501-4 and JIS B9921)

Flow Rate

1.0 CFM (28.3 L/min) ±5% accuracy (meets ISO 21501-4 and

JIS B9921)

Flow Rate Control

Electronic, automatic closed loop (patented* flow control technology)

Sample Tube Extension

Up to 39.4 ft (12 m)

Sampling Modes

Manual, automatic, beep; cumulative/differential; count or concentration

Sampling Time

1 second to 99 hours

Sampling Frequency

1 to 9999 cycles or continuous

Exhaust

Internally HEPA filtered

Vacuum Source

Internal pump

Communication Mode

Modbus® TCP over Ethernet or USB

Audible Alarm

Built-in; >85 dB at 1 meter (adjustable)

Status Indicators

Low battery, flow, laser

Alarm Output

Dry contacts, closed when alarm is engaged

Alarm Limits

Programmable for all particle channels

Display

VGA 5.7-inch (14.5-cm) touch screen display

Languages

English, German, French, Spanish, Japanese, Chinese (simplified), Italian

Software

Compatible with TrakPro Lite Secure and FMS 5 software

Printer

Built-in thermal printer (also available without printer)

Printer Output

Prints in all available languages

Environmental Sensor Interface

Supports TSI Probe Models: 960, 962, 964, 966, (air velocity/temp/RH) and Temp/RH probe

Unit ID

Configurable IP address

Security

2-level password protection to lock out usage and configuration

Data Storage

250 Zones

999 Locations

10,000 sample records including: date, time, six total particle size channels, flow status, and instrument status

Data Transfer

Via USB storage device, connection to TrakPro™ Lite Secure software through Ethernet or USB cable, or optional TSI FMS software.

Reports

Provides Pass/Fail on ISO 14644-1, EU GMP, and FS209E reports

Calibration

NIST traceable using TSI calibration system

Calibration Frequency

Recommended minimum of once per year

Dimension (H x W x D)

9.5 in. x 9.0 in. x 9.2 in. (24.1 cm x 23.0 cm x 23.2 cm) (without isokinetic inlet)

Weight

14.2 lbs (6.44 kg) with two batteries

Power

110 to 240 VAC universal power supply

Battery

Removable/rechargeable Li-Ion batteries (2 battery slots)

Battery Life

>4 hours of continuous operation

Battery Recharge

4 hours (when idle)

External Surface

Stainless steel

Operating Range

41˚to 95˚F (5˚ to 35˚C), 20% to 95% RH noncondensing

Storage Range

32˚ to 122˚F (0˚ to 50˚C), up to 98% RH noncondensing

Standards

ISO 21501-4, CE, JIS B9921

Warranty

Two years, extended warranties available

Included Accessories

Printed QuickStart guide, operating manual on CD, power supply, battery, barbed inlet, isokinetic probe, tripod, tubing, purge filter, printer paper, USB cable and TrakPro™ Lite Secure software

Optional Accessories

Electronic filter scanning probe, bas


TECHNOVALUE SOLUTIONS PVT. LTD.
X


TechnoValue Academy:
TechnoValue Solutions Pvt. Ltd.
63, Panchvati Plaza, Sector-5,
Ghansoli, Navi Mumbai-400 701.
Tel. No. : +91 22 6561 2992

Calibration Facility:

TechnoValue Solutions Pvt. Ltd.
65-A, Panchvati Plaza, Sector-5,
Ghansoli, Navi Mumbai-400 701.
Tel. No. : +91 22 6561 2992


65-A, Panchvati Plaza, Sector-5, Ghansoli, Navi Mumbai - 400701, Maharashtra, India
Ms Shravani Misal (Sales Coordinator)

Product Showcase